Inhalational anaesthesia systems — Part 7: Anaesthetic systems for use in areas with limited logistical supplies of electricity and anaesthetic gases
Primary tabs
Reference:
ISO 8835-7:2011
Publication Year:
2011
ISO 8835-7:2011 specifies safety and performance requirements for anaesthetic systems and components that:
· can function in the absence of a supply of compressed medical gases or mains electricity;
· can withstand a challenging environment, including high temperatures, humidity, shocks and vibration, and dust;
· allow local servicing and maintenance;
· may be suitable for use with flammable anaesthetic agents.